2090 CareLink Programmer

CVE-2018-10596

Medtronic 2090 Carelink Programmer Improper Restriction of Communication Channel to Intended Endpoints

Sin parche publicadounrestricted network endpoint
Descargar PNG JSON
CVSS 3.1
7.1
Alta
CVSS:3.1/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H Fuente: CNA
ExplotaciónSin evidenciaNo figura en CISA KEV
Exploits públicos0Ninguno conocido
ParcheSin parcheEl fabricante no ha publicado corrección
Exposición3.062 dDesde la divulgación, 2 jul 2018, 1:00 p. m.
Detección3.059 dDesde la divulgación hasta su ingreso al índice
Timeline

Ciclo de vida de la vulnerabilidad

Hoy
Divulgación
Publicada
Modificada
Explotación
Remediación
Monitoreo
Detectada
2 jul 18
22 may 25
17 nov 26
19 nov 26
2 jul 18
6 ago 20
10 sept 22
15 oct 24
19 nov 26
FechaFaseEventoDetalle
Divulgación Publicada CVE-2018-10596 divulgada · CVSS 7.1 (CNA)
Divulgación Modificada Última actualización del registro
Monitoreo Detectada Ingresó al índice
Resumen

Medtronic 2090 CareLink Programmer uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.

Vector CVSS 3.1

Desglose de métricas

Opción resaltada = valor del vector · izquierda = peor caso · CNA
Vector de ataqueAV:N
RedAdyacenteLocalFísica
ComplejidadAC:H
BajaAlta
PrivilegiosPR:L
NingunoBajosAltos
InteracciónUI:R
NingunaRequerida
AlcanceS:U
CambiadoSin cambio
ConfidencialidadC:H
AltoBajoNinguno
IntegridadI:H
AltoBajoNinguno
DisponibilidadA:H
AltoBajoNinguno
Versiones

Rangos afectados

Afectada Corregida
2090 CareLink Programmer · 29901 Encore Programmertodas
Remediación

Estado del parche por paquete

Paquete / productoEstadoCorregida enDesde divulgaciónÚltima versión
2090 CareLink Programmer
Sin parche — — —
29901 Encore Programmer
Sin parche — — —
Explotación

Arsenal público

Sin exploits, módulos ni PoC públicos conocidos a la fecha de verificación.

Registro
EscenarioGENERALicscert
MitigaciónMedtronic has assessed the vulnerabilities and determined that no new potential safety risks were identified. In order to enhance system security, Medtronic has added periodic integrity checks for files associated with the software deployment network. Additionally, Medtronic has developed server-side security changes that further enhance security. Medtronic reports that they will not be issuing a product update; however, Medtronic has identified compensating controls within this advisory to reduce the risk of exploitation and reiterates the following from the CareLink 2090 Programmer Reference Manual: * Maintain good physical controls over the programmer. Having a secure physical environment prevents access to the internals of the programmer. * Only connect the programmer to managed, secure networks. * Update the software on the programmer when Medtronic updates are available. * Alternatively, disconnect the programmer from the network. Network connectivity is not required for normal programmer operation. * Offline updates are available, contact your Medtronic representative for more information. Medtronic has deployed mitigating patches to address the reported vulnerabilities. Medtronic has also stated that they have increased security controls associated with these vulnerabilities. As a result of the available mitigating patches, Medtronic has re-enabled the network-based software update mechanism. Medtronic has stated that the patch for affected products can be obtained by contacting Medtronic Technical Services at 800‑638‑1991. After additional review and risk evaluation of the affected products, Medtronic has disabled the network-based software update mechanism, including both the VPN and the HTTP subservices, as an immediate security mitigation. Users should not attempt to update the affected products over the network as this update mechanism is vulnerable to the attack described in section 4.2.3. Medtronic will continue to implement and deploy increased security protections and mitigations to address the vulnerabilities in this advisory. Users should still obtain and apply updates via controlled USB dongles and should contact their Medtronic representative for more information. Medtronic recommends that affected products continue to be used for their intended purpose in the previously described manner. Medtronic has released a security bulletin for the 2090 CareLink Programmer https://www.medtronic.com/security .icscert
Referencias
Fabricantehttps://global.medtronic.com/xg-en/product-security/security-bulletins/carelink-2090-29901.html
Otrahttps://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-058-01
Otrahttps://ics-cert.us-cert.gov/advisories/ICSMA-18-058-01