CVE-2018-14781
Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Authentication Bypass by Capture-replay
CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N
Fuente: CNA
Ciclo de vida de la vulnerabilidad
| Fecha | Fase | Evento | Detalle |
|---|---|---|---|
| Divulgación | Publicada | CVE-2018-14781 divulgada · CVSS 5.3 (CNA) | |
| Divulgación | Modificada | Última actualización del registro | |
| Monitoreo | Detectada | Ingresó al índice |
Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
Desglose de métricas
Rangos afectados
Estado del parche por paquete
| Paquete / producto | Estado | Corregida en | Desde divulgación | Última versión |
|---|---|---|---|---|
MMT- 508 - MiniMed pump |
Sin parche | — | — | — |
MMT – 511 pump Paradigm |
Sin parche | — | — | — |
MMT – 512 / MMT – 712 Paradigm x12 |
Sin parche | — | — | — |
MMT – 515 / MMT – 715 Paradigm x15 |
Sin parche | — | — | — |
MMT – 522 / MMT – 722 Paradigm REAL-TIME |
Sin parche | — | — | — |
MMT – 522(K) / MMT – 722(K) Paradigm REAL-TIME |
Sin parche | — | — | — |
MMT – 523 / MMT – 723 Paradigm Revel |
Sin parche | — | — | — |
MMT – 523(K) / MMT – 723(K) Paradigm |
Sin parche | — | — | — |
MMT – 554 / MMT – 754 MiniMed Veo |
Sin parche | — | — | — |
MMT – 551 / MMT – 751 MiniMed 530G |
Sin parche | — | — | — |
Arsenal público
Sin exploits, módulos ni PoC públicos conocidos a la fecha de verificación.
https://global.medtronic.com/xg-en/product-security/security-bulletins/minimed.htmlhttps://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02http://www.securityfocus.com/bid/105044Las referencias externas se muestran para copiar: cve.llama.ec no lo redirige a otros sitios.