CVE-2019-13539
Medtronic Valleylab FT10 and FX8 Reversible One-way Hash
CVSS:3.1/AV:L/AC:H/PR:L/UI:N/S:U/C:H/I:H/A:H
Fuente: CNA
Ciclo de vida de la vulnerabilidad
| Fecha | Fase | Evento | Detalle |
|---|---|---|---|
| Divulgación | Publicada | CVE-2019-13539 divulgada · CVSS 7 (CNA) | |
| Divulgación | Modificada | Última actualización del registro | |
| Monitoreo | Detectada | Ingresó al índice |
Medtronic Valleylab Exchange Client version 3.4 and below, Valleylab FT10 Energy Platform (VLFT10GEN) software version 4.0.0 and below, and Valleylab FX8 Energy Platform (VLFX8GEN) software version 1.1.0 and below use the descrypt algorithm for OS password hashing. While interactive, network-based logons are disabled, and attackers can use the other vulnerabilities within this report to obtain local shell access and access these hashes.
Desglose de métricas
Rangos afectados
Estado del parche por paquete
| Paquete / producto | Estado | Corregida en | Desde divulgación | Última versión |
|---|---|---|---|---|
Valleylab Exchange Client |
Sin parche | — | — | — |
Valleylab FT10 Energy Platform (VLFT10GEN) |
Sin parche | — | — | — |
Valleylab FX8 Energy Platform (VLFX8GEN) |
Sin parche | — | — | — |
Arsenal público
Sin exploits, módulos ni PoC públicos conocidos a la fecha de verificación.
https://global.medtronic.com/xg-en/product-security/security-bulletins/valleylab-generator-rfid-vulnerabilities.htmlhttps://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-311-02https://www.us-cert.gov/ics/advisories/icsma-19-311-02Las referencias externas se muestran para copiar: cve.llama.ec no lo redirige a otros sitios.