CVE-2019-6538
Medtronic Conexus Radio Frequency Telemetry Protocol Improper Access Control
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H
Fuente: CNA
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:NCiclo de vida de la vulnerabilidad
| Fecha | Fase | Evento | Detalle |
|---|---|---|---|
| Divulgación | Publicada | CVE-2019-6538 divulgada · CVSS 9.3 (CNA) | |
| Divulgación | Modificada | Última actualización del registro | |
| Monitoreo | Detectada | Ingresó al índice | |
| Monitoreo | Verificada | Última verificación contra las fuentes |
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.
Desglose de métricas
Rangos afectados
Estado del parche por paquete
| Paquete / producto | Estado | Corregida en | Desde divulgación | Última versión |
|---|---|---|---|---|
Conexus Radio Frequency Telemetry Protocol |
Sin parche | — | — | — |
MyCareLink Monitor |
Sin parche | — | — | — |
CareLink Monitor |
Sin parche | — | — | — |
CareLink 2090 Programmer |
Sin parche | — | — | — |
Amplia CRT-D |
Sin parche | — | — | — |
Claria CRT-D |
Sin parche | — | — | — |
Compia CRT-D |
Sin parche | — | — | — |
Concerto CRT-D |
Sin parche | — | — | — |
Concerto II CRT-D |
Sin parche | — | — | — |
Consulta CRT-D |
Sin parche | — | — | — |
Evera ICD |
Sin parche | — | — | — |
Maximo II CRT-D |
Sin parche | — | — | — |
Maximo II ICD |
Sin parche | — | — | — |
Mirro ICD |
Sin parche | — | — | — |
Nayamed ND ICD |
Sin parche | — | — | — |
Primo ICD |
Sin parche | — | — | — |
Protecta ICD, Protecta CRT-D |
Sin parche | — | — | — |
Secura ICD |
Sin parche | — | — | — |
Virtuoso ICD |
Sin parche | — | — | — |
Virtuoso II ICD |
Sin parche | — | — | — |
Visia AF ICD |
Sin parche | — | — | — |
Viva CRT-D |
Sin parche | — | — | — |
Brava CRT-D |
Sin parche | — | — | — |
Mirro MRI ICD |
Sin parche | — | — | — |
Arsenal público
Sin exploits, módulos ni PoC públicos conocidos a la fecha de verificación.
https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01http://www.securityfocus.com/bid/107544Las referencias externas se muestran para copiar: cve.llama.ec no lo redirige a otros sitios.