llamaCVEVulnerability Intelligence
/
Smart Model 25000 Patient Reader

CVE-2020-27252

Medtronic MyCareLink Smart Time-of-check Time-of-use Race Condition

Sin parche publicadorace condition
Descargar PNG JSON
CVSS 3.1
8.8
Alta
CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:H/A:H Fuente: CNA
ExplotaciónSin evidenciaNo figura en CISA KEV
Exploits públicos0Ninguno conocido
ParcheSin parcheEl fabricante no ha publicado corrección
Exposición2.165 dDesde la divulgación, 14 dic 2020, 2:19 p. m.
Detección2.163 dDesde la divulgación hasta su ingreso al índice
Timeline

Ciclo de vida de la vulnerabilidad

Hoy
Divulgación
Publicada
Modificada
Explotación
Remediación
Monitoreo
Detectada
14 dic 20
22 may 25
17 nov 26
19 nov 26
14 dic 20
9 jun 22
2 dic 23
27 may 25
19 nov 26
FechaFaseEventoDetalle
Divulgación Publicada CVE-2020-27252 divulgada · CVSS 8.8 (CNA)
Divulgación Modificada Última actualización del registro
Monitoreo Detectada Ingresó al índice
Resumen

Medtronic MyCareLink Smart 25000 is vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited, an attacker could remotely execute code on the MCL Smart Patient Reader device, leading to control of the device.

Vector CVSS 3.1

Desglose de métricas

Opción resaltada = valor del vector · izquierda = peor caso · CNA
Vector de ataqueAV:A
RedAdyacenteLocalFísica
ComplejidadAC:L
BajaAlta
PrivilegiosPR:N
NingunoBajosAltos
InteracciónUI:R
NingunaRequerida
AlcanceS:C
CambiadoSin cambio
ConfidencialidadC:H
AltoBajoNinguno
IntegridadI:H
AltoBajoNinguno
DisponibilidadA:H
AltoBajoNinguno
Versiones

Rangos afectados

Afectada Corregida
Smart Model 25000 Patient Readertodas
Remediación

Estado del parche por paquete

Paquete / productoEstadoCorregida enDesde divulgaciónÚltima versión
Smart Model 25000 Patient Reader
Sin parche — — —
Explotación

Arsenal público

Sin exploits, módulos ni PoC públicos conocidos a la fecha de verificación.

Registro
EscenarioGENERALicscert
MitigaciónIn response to these vulnerabilities, Medtronic has applied additional controls for monitoring and responding to improper use of the MCL Smart Patient Reader: * Medtronic has implemented enhanced integrity validation (EIV) technology, which provides early detection and real-time mitigation of known vulnerability exploitation attempts. * Medtronic has also implemented advanced detection system technology, which enables device-level logging and monitoring of all device activity and behavior. Medtronic recommends that users take additional defensive measures to minimize risk. Specifically, users should: * Maintain good physical control over home monitors. * This includes only using home monitors in private environments such as a home, apartment, or otherwise physically controlled environment. * Use only home monitors obtained directly from your healthcare provider or a Medtronic representative. * Patients should ensure that the operating system of their mobile phone is updated to the latest version of the available Android or Apple iOS operating system. Report any concerning behavior regarding these products to your healthcare provider or a Medtronic representative.icscert
solutionA firmware update to eliminates these vulnerabilities has been developed by Medtronic and is available by updating the MyCareLink Smartapp via the associated mobile application store. Upgrading to the latest v5.2 mobile application version will ensure the Patient Reader is also updated on next use. The user’s smart phone must be updated to the following operating system version for the patches to be applied: iOS 10 and above; Android 6.0 and above. Medtronic has released additional patient focused information https://www.medtronic.com/security : https://www.medtronic.com/xg-en/product-security/security-bulletins.html https://www.medtronic.com/xg-en/product-security/security-bulletins.htmlicscert
Referencias
Fabricantehttps://global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-smart-security-vulnerability-patch.html
Otrahttps://www.cisa.gov/news-events/ics-medical-advisories/icsma-20-345-01
Otrahttps://us-cert.cisa.gov/ics/advisories/icsma-20-345-01

Las referencias externas se muestran para copiar: cve.llama.ec no lo redirige a otros sitios.